pH Measurement Best Practices: Calibration, Electrode Care and Choosing a Meter
By Ramo Trading··3 min read
pH affects chemical reactions, corrosion, disinfection, biological treatment and product quality. That is why pH is measured in nearly every lab and treatment plant. It is also easy to get wrong. Most bad pH readings come down to calibration or electrode care, not the meter itself.
1. Calibrate the right way
Use at least two buffers that bracket your samples. For example, use pH 4 and 7 for acidic samples, or 7 and 10 for alkaline samples. Three-point calibration (4, 7, 10) covers a wide range.
Use fresh buffer. Pour a small amount into a clean container and never pour it back into the bottle. Color-coded red and yellow calibration beakers help prevent mix-ups.
Match temperatures. Use automatic temperature compensation and let buffers and samples reach the same temperature.
Check slope and offset. A slope well below the ideal range is a sign that the electrode is aging and needs cleaning or replacing.
Calibrate often. Daily is typical for routine work. Calibrate before every batch for critical measurements.
2. Look after the electrode
Hach IntelliCAL PHC725 Red Rod pH electrode
Keep the glass bulb hydrated. Store it in electrode storage solution, never in deionized water, which pulls ions out of the glass.
Rinse between samples with deionized water and blot gently. Do not wipe the bulb.
Clean for the type of contamination: mild detergent for oils, dilute acid for mineral deposits, and enzyme cleaners for proteins.
Refill electrolyte on refillable electrodes, and open the fill hole while measuring.
Replace it when it is time. Electrodes are consumables, and a slow, drifting electrode costs more in lost time than a new one.
For field data you can document, choose the Hach HQ1110 portable pH/ORP meter or the multi-parameter Hach HQ2200 for pH, conductivity and TDS. It is also available with a gel pH probe. Hach's IntelliCAL digital probes store their own calibration data, so you can swap them between meters.
Benchtop meters
For labs, the Hach HQ430d benchtop meter offers a stable reading platform and supports good laboratory practice.
A practical overview of the leading water-testing brands, what each does best, and how to build the right testing setup for drinking water, wastewater, industrial water,...
Ramo Trading has published a series of practical buying and application guides for water and wastewater testing, covering Hach, CHEMetrics, LaMotte, Palintest, IDEXX and more.
Total coliforms and E. coli are the main indicators of microbiological water safety. Compare the two most common test methods, IDEXX Colilert enzyme substrate testing and...
Comments (0)