| Column Oven: | |
| Temperature Range | Ambient temperature +7°C ~ 400°C (in 1°C increment) |
| Temperature Control Accuracy | ± 0.02°C |
| Programmed temperature setting | 0.1°C ~ 40°C/min (in 1°C increment) |
| Program ramps | 7 ramps in total (10 ramps available with control workstation) |
| Cooling time | 400°C to 50°C in 8-10 min at 25°C ambient |
| Size (LxWxH) | 284x280x241mm (internal) 340x345x281mm (external) |
| Hydrogen flame ionization detector (FID): | |
| Detection limit | ≤ 3x10-12 g/s (C16) |
| Best test result | ≤ 3x10-12 g/s (C16) |
| Baseline noise | ≤ 5x10-14 A |
| Baseline drift | ≤ 6x10-13 A /30 min |
| Linear range | ≥ 106 |
| Thermal Conductivity Detector (TCD): | |
| Sensitivity | ≥5000 mV.ml / mg (C 16) |
| Baseline noise | ≤20 μV |
| Baseline drift | ≤60 μV/h |
| Linearity range | ≥104 |
| Flame Photometric Detector (FPD): | |
| Detection limit | ≤8x10 g / s (P) |
| ≤8x10 g / s (S) | Flame Photometric Detector (FPD): Drift |
| ≤2x10-11 A/30 min | Flame Photometric Detector (FPD): Baseline noise |