SUBSCRIBE TO OUR NEWSLETTER
Register now to get updates on promotions and coupons.


Estimated Delivery 10-14 business day
Couldn't load pickup availability
Free standard shipping on orders over 185 AED
| Ion Chromatographic Pump: | |
| Pressure Display Accuracy | ≤ 0.1 MPa |
| Maximum Pressure | 35 Mpa (PEEK) |
| Flow Range | 0.001 ~ 9.999 mL/min |
| Resolution of Flow Rate | 0.001 ml |
| Flow Precision | < 0.1% |
| Flow Accuracy | < 0.1% |
| Pressure Pulse | ≤ 0.5% |
| Flow Stability | (0.2-0.5) mL/min ≤ 3%; (0.5-1.0) mL/min ≤ 2%; > 1.0 mL/min ≤ 2% |
| Numerical-control and Electromagnetic Sample Injector: | |
| Maximum Pressure | 35 Mpa |
| Power Supply | 24 V (DC) |
| Column Heater: | |
| Operating Temperature Range | +20°C~60°C (68~140°F) |
| Controlling Temperature Accuracy | ± 0.01°C |
| Allowable Deviation of Column Heater’s Temperature | ± 1°C |
| Temperature Stability | ≤ 0.05°C/h |
| Conduction Detection System: | |
| Cell Volume | ≤0.8μL |
| Detection Mode | Bipolar conductivity detection |
| Detection Range | 0~50000 μS/cm |
| Detection Resolution | ≤0.0020nS/cm |
| Output Voltage | -6000~+6000 mv (adjustable) |
| Electronic Noise | 0.02 nS |
| Baseline Noise | ≤ 0.001 μS/cm |
| Baseline Drift | ≤ 0.02μS |
| Operating Temperature Range | Room temperature +5°C~60°C(41~140°F) |
| Controlling Temperature Accuracy | ±0.01°C |
| Maximum Pressure | 10.0 Mpa |
| Linear Range | ≥ 10 3 |
| Instrument Linearity | ≥0.999 |
| Quantitative Repeatability | ≤1.0% |
| Qualitative Repeatability | ≤1.0% |
| Thermal Buffer System of Eluent: | |
| Thermal Buffer System of Eluent | Before enter into the column,the eluent is preheated. By the way,can avoid the rapid heating up and the bubbles to generate,the baseline is more stable, effectively shorten the start-up balance time and improve the analysis efficiency and effect. |
| Temperature Range | 25~40°C (77~104°F) |
| Built-in and Low-pressure Degassing Device: | |
| Vacuum Degree | -70 kPa |
| Maximum Flow Rate | 10 mL/min |
| Internal Volume | 30 μL |
| Degassing Efficiency | 10 mL/min 90% |
| Flow System: | |
| Six-way Valve | PEEK material, pressure 5000 psi; Independent automatic collecting and flow function. |
| Built in Eluent Generator: | |
| Eluent Types | KOH/MSA |
| Eluent Concentration Range | 0.1-120 mM |
| Concentration Increment | 0.1 mM |
| Flow Rate Range | 0.1-5.0 mL/min |
| Maximum Pressure | 20 Mpa |
| Minimum Pressure | 5 Mpa |
| Suppressor: | |
| Type | Self-Regenerating electrolytic micro-membrane suppressor |
| Maximum Pressure | 6.0 Mpa |
| Dead Volume | <50 μL |
| Other Specifications: | |
| Dimension (LxWxH) | 500x500x760 mm |
| Net Weight | 48 kg |
| Gross Weight | 73 kg |
| Power | 350 W |
Download Catalog