| Maximum capacity [Max] | 82 / 220 g |
| Minimum load | 1 mg |
| Readability [d] | 0.01 / 0.1 mg |
| Verification unit [e] | 1 mg |
| Tare range | -220 g |
| Standard repeatability [5% Max] | 0.005 mg |
| Standard minimum weight (USP) | 10 mg |
| Standard minimum weight (U=1%, k=2) | 1 mg |
| Permissible repeatability [5% Max] | 0.012 mg |
| Permissible repeatability [Max] | 0.08 mg |
| Linearity | ±0.06 / 0.2 mg |
| Eccentric load deviation | 0.2 mg |
| Sensitivity time drift | 1×10⁻⁶/Year×Rt |
| Stabilization time | 4 s |
| Adjustment | internal (automatic) |
| OIML Class | I |
| Leveling system | semi-automatic – LevelSENSING |
| Display | 10″ graphic colour touchscreen |
| Weighing chamber doors | manual |
| Delivery components | Analytical Balance, weighing pan, weighing pan shield, centring ring, bottom cover, brush, fabric dust cover, power supply. |
| Weighing chamber dimensions | 168×160×228 mm |
| Weighing pan dimensions | ø90 open-work pan + ø85 (option) mm |
| Packaging dimensions W x D x H | 750×492×595 mm |
| Net weight | 9.8 kg |
| Gross weight | 16 kg |
| Protection class | IP 43 |
| Communication interface | 2×USB-A, USB-C, RS 232 (COM3), HDMI, Ethernet, Wi-Fi®, Hotspot |
| Power supply | Adapter: 100 – 240V AC 50/60Hz 1A; 15V DC 2.4A |
| Operating temperature | +10 – +40 °C |
| Operating temperature change rate | ±0.3 °C / 1 h (±1 °C / 8 h) |
| Relative humidity | 20% – 80% |
| Relative humidity change rate | ±1% / h (±4% / 8 h) |















Data Sheet
User Manual
Software Manual
Software Manual - CBCP-07
User Manual - 5Y Series Balances
Start up Guide
Software Manual
Conformity Declarations
Laboratory balance cleaning
Pharmacy - Mass Measurement Guide
RTP Thermal Receipt Printer
USP New Regulation
Lexicon of Basic Metrological Terms
RFID
Hybrid Protective Glass
Laboratory Balances
Metrology in the laboratory 2024
SQC in laboratory
Checking and calibrating fixed volume and adjustable volume piston pipettes using a gravimetric method
XA 5Y Micro and Analytical Balances
Laboratory Balances