Dissolved oxygen (DO) tells you whether a body of water can support life and whether a treatment process is working. In wastewater aeration basins, aquaculture tanks, rivers and brewery process water, a wrong DO reading means wasted aeration energy, stressed organisms or a failed compliance sample. The choice of sensor and how you look after it matter as much as the meter itself.
Two sensor technologies
| Feature |
Optical (LDO) |
Clark-type (membrane) |
| How it measures |
A sensing layer glows; oxygen shortens that glow, and the meter reads the change. |
Oxygen crosses a membrane and is reduced at an electrode, producing a small current. |
| Consumes oxygen |
No |
Yes, so the sample needs movement |
| Upkeep |
Replace the sensor cap periodically |
Replace membrane and electrolyte, polish electrodes |
| Warm-up |
Short |
Needs polarisation time (polarographic types) |
| Best for |
Field work, long deployments, low-DO samples |
Budget lab spot checks where staff are trained |
For most new purchases we recommend optical LDO. There is no membrane to puncture, no electrolyte to refill and no stirring requirement, which removes the most common causes of bad DO readings.
What changes a DO reading
-
Temperature: colder water holds more oxygen. Your meter should compensate automatically.
-
Salinity: saltier water holds less. Enter salinity for brackish or seawater samples.
-
Barometric pressure and altitude: they change the saturation value used for calibration.
-
Sensor fouling: biofilm, oil or scale slows the response and causes drift.
Reporting tip: state whether a result is mg/L or % saturation. They are not interchangeable, and permits normally specify one.
Calibration in practice
- Rinse the probe with deionised water and blot dry. Do not rub the sensing surface.
- Use the manufacturer's air-saturated water or water-saturated air method and let the reading stabilise.
- Confirm temperature and pressure are correct, then accept the calibration.
- Check against a second reference now and then. A Winkler titration is the classic reference method.
- Record the calibration slope or status so you can see drift over time.
Hach DO equipment we supply
FAQ
Why does my DO reading drift low in still water?
A Clark-type sensor consumes oxygen at the membrane, so in still water it depletes the layer around it. Stir the sample or move the probe. Optical sensors do not have this problem.
How often should I replace an LDO sensor cap?
Follow the manufacturer's schedule, which depends on the model and conditions. Replace it sooner if calibration fails or the reading drifts.
Need help choosing a DO meter?
Tell us your application and we will recommend a meter, probe and cable length.