| Extraction Method | Magnetic Bead Method |
| Working Mode | Automatic sampling + Nucleic acid extraction + PCR reaction system addition |
| Throughput | 1-96, Linear slide type sample rack |
| Extraction Volume | 20-1000 ul |
| Processing Time | Complete the processing of 96 samples within 60 minutes (related to reagents) |
| Magnetic Bead Recovery | ≥98% |
| Temp Range | RT-105℃, Lysis and elution position |
| Temp Accuracy | 0.1℃ |
| Heating Method | Dry bath heating |
| Heating Speed | RT-100℃≤6min |
| Shaking Function | Up and down oscillation (1-5 gears adjustable) |
| Extraction Position | 6 (96-well deep well plate) |
| Robotic arm | A robotic arm for adding samples and reagents |
| Pipetting Channel | 2 Channel |
| Liquid Detection | Pneumatic liquid level detection principle, intelligent detection of blocked needle |
| Pipetting Tip | 50ul,200ul,1000ul, Disposable black conductive needle with filter element |
| Tip Amount | 2-3 Tips/sample |
| Pipetting Accuracy | 10ul, CV≤3.0%, Accuracy≤5.0%, 50ul Tip 50ul, CV≤2.0%, Accuracy≤2.0%, 1000ul Tip 100ul, CV≤1.5%, Accuracy≤2.0%, 1000ul Tip |
| Sample Volume | 2-1000 ul |
| Working Zone | 2 PCR positions with cooling function 6 Tip positions for three types of Tips 2 Reagent positions (5ml freezing tube rack position with cooling function, one reserved position) |
| Protective function | Start up self-test, Power-off protection, High temperature alarm, Over-temperature protection, Tip removal protection |
| Disinfection method | UV lamp (30Wx1, 8xW1) |
| Illumination Lamp | 10W LED lamp |
| Audible Alarm | Yes (Red and blue blinking) |
| Safety Door Design | With safety lock function, the safety door is opened and the program is suspended |
| Display | 10.1inch touch screen, Windows System |
| Scanning | Optional |
| Interface | LAN interface (Bi-direction LIS optional) |
| Contamination control | Built-in air duct and HEPA filter can effectively filter internal aerosols and prevent cross-contamination |
| IAP Function | Firmware can be upgraded online at any time |
| External Size | 1420x850x1842 mm |
| Package Size | 1535x970x1180 mm (Main instrument) 1540x970x1160 mm(Base) |
| Gross Weight | 360kg(Main instrument) 190kg(Base) |
Download Catalog